Specification
OPTICAL SPECIFICATIONS
- Light Source: LEDs source with 8 Wavebands
- Nominal Wavebands: For Tests performed from 340nm, 405nm, 450nm, 505nm, 545nm, 580nm, 630nm, 670nm. Optional additional waveband of 700mm
- Optical System: Semi Micro Cuvette
- Reagent Volume: Lowest Reagent Volume of 250 µl ( 0.250ml)
PROGRAMMING
- Measurement Modes: Kinetic fixed time end point multi standard calibration monochromatic&Bi-chromatic absorbance & scan capacity
- Memory : 100 User defined Programs, 1000 Results Program Memory
Input /Output
- Display: 5 Inch Graphic LCD (240 X 128 px)
- Keyboard: Spillproof Membrane Keyboard
- Thermal Printer :57mm paper Easy Load with Quick Ac
- Output Interface : Serial Port (RS 232 ) Output Direct t connectivity to external printer and computer
- Self Testing : Automatic self testing of entire optical system for all wavelengths in config mode.
THERMOSTAT SYSTEM
- Temperature: RT & 37°C Temperatures with + / 0.5 ºC. Precision
- Thermostat: Digitally Controlled Solid State Sensor and Heater
POWER, BATTERY BACKUP & DIMENSIONS
- Power Supply : POWER ADAPTOR 75W, AC Mains: 180 V to 240 V AC, 50 Hz
- Battery backup: DC: 10 V to 18 V DC 1A, ( 12V Motorcycle Battery runs for 8 hrs
- Operating Req.: Room Temperature: 10°C to 35 °C. Protect from Direct Sunlight.
- Dimensions: L:400mm x W:300mm x 140mm Appx.
- Weight: 2.0 kg Appx & 5.5Kg Approx along with Packaging